In the Service functions screen you can access a number of screens:
Device information
Statistics
Inputs
Outputs
Voltage and temperature monitor
Functional tests
Adjustment and calibration
Service information can be used in cooperation with Struers Service for remote diagnostics of the
equipment.
12.8 Disposal
Equipment marked with a WEEE symbol contains electrical and electronic
components and must not be disposed of as general waste.
Contact your local authorities for information on the correct method of disposal in
accordance with national legislation.
For disposal of consumables and recirculation fluid, follow local regulations.
13 Troubleshooting - LaboPol-60
13.1 Troubleshooting - LaboPol-60
Error Cause Action
Noise when the machine
starts, or the turntable will not
turn.
The belt is not tight enough. The belt must be tightened.
Contact Struers Service.
The machine does not
operate when the start switch
is pressed.
The main switch is off. Turn the main switch on.
The fuse is blown (located at
the back of the machine).
Replace the fuse.
Water is not draining away. Drain hose squeezed. Straighten the hose.
Drain hose clogged. Clean the hose.
Drain hose does not slope
downwards.
Adjust the hose to an even slope.
86 LaboPol-60
13 Troubleshooting - LaboPol-60