12.5 Annually
12.5.1 Testing safety devices
The safety devices must be tested at least once a year.
WARNING
Do not use the machine with defective safety devices.
Contact Struers Service.
Note
Testing should always be performed by a qualified technician (electromechanical,
electronic, mechanical, pneumatic, etc.).
12.5.2 Emergency stop
Test 1
1. Press the Start button. The machine starts operating.
2. Press the emergency stop.
3. If operation does not stop, press the Stop button.
4. Contact Struers Service.
Test 2
1. Press the emergency stop.
2. Press the Start button.
3. If the machine starts, press the Stop button.
4. Contact Struers Service.
12.6 Spare parts
Technical questions and spare parts
If you have technical questions or when you order spare parts, state the serial number. The serial
number is stated on the type plate of the unit.
For further information, or to check the availability of spare parts, contact Struers Service. Contact
information is available on Struers.com.
84 LaboPol-60
12 Maintenance and service - LaboPol-60