7.2.6 Adjusting the force
Hint
Do not use maximum force at the same time as maximum speed.
1. To adjust the force, turn the
force adjustment screw.
The indications on the
housing correspond to the
actual force in Newton.
A Force indicator
B Force
adjustment
screw
C Pressure foot
Indication Force
(0) 0 - 5 N
1 10 N
2 20 N
3 30 N
7.2.7 Manual preparation
If you cannot prepare a specimen by using a standard specimen mover plate or specimen holder, you
can prepare it manually.
When you perform manual preparation, you hold the specimen in your hand and press it firmly onto
and across the preparation surface.