# Error Message Explanation Action
26 Information
Specimen mover plate not
lowered!
The pneumatic head with
specimen mover plate is
not down when the
process is started.
Appears if method is
started in single specimen
mode (SS) and mover
plate is not lowered.
Lower the pneumatic
head.
Re-start the process.
13.4 LaboForce-Mi
Error Cause Action
Specimen holder plate
vibrates.
Specimen holder plate screws
loose.
Tighten specimen holder
plate screws.
Specimen holder plate
unbalanced.
Replace specimen holder
plate.
The preparation disc runs
unevenly or stops.
Force too high. Reduce the force.
The preparation disc stops. Frequency inverter has
stopped the equipment.
Switch the equipment off.
Wait for a few minutes, then
re-start.
If the error remains, contact
Struers Service.
The column starts to turn. Column screws loose. Tighten the screws
immediately.
The force adjustment screw
rotates by itself when a
specimen passes the centre
of the preparation disc.
The friction between the force
adjustment screw and the
rubber pressure foot is too
high.
Add a drop of oil on the
contact surface of the rubber
pressure foot to reduce the
friction.
Uneven specimens. Specimens are passing over
the center of the disc.
Reposition the horizontal
position of the control panel.
13 Troubleshooting - LaboPol-60
LaboPol-60 91