Subject Specifications
Electrical power supply
cables
Voltage/frequency Local standards can override the
recommendations for the main
electrical power supply cable. If
needed, contact a qualified
electrician to verify which option
is suitable for the local installation
setup.
3 x 200-240 V/50-60 Hz • Min. fuse:
25 A
• Minimum cable size at
minimum fuse:
3 x AWG12/2.5 mm
2
+ PE
• Max. fuse:
35 A
• Minimum cable size at
maximum fuse:
3 x AWG12/2.5 mm
2
+ PE
3 x 380-480 V/50-60 Hz • Min. fuse:
15 A
• Minimum cable size at
minimum fuse:
3 x AWG12/2.5 mm
2
+ PE
• Max. fuse:
35 A
• Minimum cable size at
maximum fuse:
3 x AWG12/2.5 mm
2
+ PE
Residual Current Circuit
Breaker (RCCB)
Type B,30mA is required.
Note
Always follow local regulations.
Software and electronics Display LCD 320 x 240 pixels
Controls Touchpad/Turn/Push knob
Memory FLASH-ROM/RAM/NV-RAM
9 Technical data
AbraPol-30 55