11.3 Noise and vibration levels
Noise level A-weighted sound
emission pressure
level at workstations
L
pA
= 63 dB(A) (measured value)
Uncertainty K = 4 dB
Measurements made in accordance with EN
ISO 11202
Vibration level During preparation Total vibration exposure to upper parts of the
body does not exceed 2.5 m/s
2
.
11.4 Technical data - equipment units
For technical data covering the individual units of equipment, see the specific equipment manual.
11.5 Safety Related Parts of the Control System (SRP/CS)
WARNING
Safety critical components must be replaced after a maximum lifetime of 20 years.
Contact Struers Service.
Note
SRP/CS (safety-related parts of a control system) are parts that have an influence
on safe operation of the machine.
Note
Replacement of safety critical components must only be performed by a Struers
engineer or a qualified technician (electromechanical, electronic, mechanical,
pneumatic, etc.).
Safety critical components must only be replaced by components with at least the
same safety level.
Contact Struers Service.
Safety
related part
Manufacturer/Manufacturer
description
Manufacturer
catalog no.
Electrical
ref.
Struers
catalog no.
Emergency
stop button
Schlegel
Latching mushroom head
ES Ø22 type
RV
S1 2SA10400
Emergency
stop contact
Schlegel
Modular contact, momentary
1 NC type MTO S1 2SB10071
Module
holder
Schlegel
Module holder. 3 elem. MHR-3
MHR-3 S1 2SA41603
11 Technical data
LaboPol-20 47