Stryker 1089 - Preventative Maintenance

Stryker 1089
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Return To Table of Contents
14 1089-009-002 REV B www.stryker.com
All fasteners secure
Siderails move and latch properly
Engage brake pedal and push on the stretcher to ensure all casters lock securely
Steer function working properly
All casters secure and swivel properly
Inspect each caster and remove any wax or debris which may have collected on the caster or braking
mechanism
Body restraints working properly
I.V. pole intact and operating properly
Oxygen bottle holder intact and operating properly
Fowler operating and latching properly
Optional articulating head piece locking and releasing properly
Gatch operating properly
Trendelenburg/reverse Trendelenburg operating properly
No rips or cracks in mattress cover
Arm boards intact and operating properly
Arm board support levers intact and operating properly
Ground chain intact
No leaks at hydraulic connections
Hydraulic jacks holding properly
Hydraulic drop rate set properly
Hydraulic oil level sufficient
Lubricate where required (page 15)
Accessories and mounting hardware in good condition and working properly
At a minimum, preventative maintenance should be performed annually. A preventative maintenance program should
be established for all Stryker Medical equipment. Preventative maintenance may need to be performed more frequently
based on the usage level of the product.
Preventative Maintenance
Product Serial Number:
Completed by: ____________________________________________ Date: _____________________