Return To Table of Contents
32 1089-009-002 REV B www.stryker.com
Service Information
PNEUMATIC FOWLER RELEASE HANDLE ADJUSTMENT
Tools Required:
• #2 Phillips Screwdriver
Procedure:
1. Apply the brakes.
2. Using a #2 Phillips screwdriver, loosen the two screws on the handle linkage. Release the handle tension spring
on the patient left side.
3. Adjust the handles (B), so there is a 3/4” gap between the inner fowler frame and the handle.
4. Tighten the two screws on the handle linkage bar and reattach the spring.
5. Check the adjustment of the pneumatic cylinders. See “Pneumatic Fowler Adjustment” on page 31.
6. Verify proper operation before returning the unit to service.