Setup Procedures
Make sure that the unit is working properly before it is put into service. The following list will ensure that each part of
the unit is checked.
1. Fully depress the brake pedal of the stretcher to set the four wheel brakes and verify that all four casters are
locked.
2. Raise and lower the hydraulic lift system.
3. Raise the unit completely and activate the Trendelenburg function. Ensure that the head end lowers to the lowest
position.
4. Raise the unit completely and activate the reverse Trendelenburg function. Ensure that the foot end lowers to the
lowest position.
5. Activate the fifth wheel to ensure that it is operating properly.
6. Ensure that the siderails raise and lower smoothly and lock securely when fully raised.
7. Raise and lower the fowler (head end).
8. Rotate the head piece in both directions to ensure that it is operating properly.
CAUTION
Do not modify this stretcher. Modifying the unit can cause unpredictable operation resulting in injury to the patient or
operator. Modifying the unit will also void its warranty.