www.stryker.com 1089-009-002 REV B 27
Service Information
LITTER REMOVAL
Tools Required:
• 1/2” Socket
• 3/8” Drive Ratchet
• 7/16” - 20 Fine Thread Bolt
Procedure:
WARNING
Two people are required to safely perform this procedure.
1. Apply the brakes.
2. Raise the litter to its highest height.
3. Remove the mattress.
4. Using a 1/2” socket and 3/8” drive ratchet, remove the two screws that hold the jack support tubes to the jack
shafts.
5. Thread a 7/16” - 20 fine thread bolt far enough into the top of the jack support tubes to separate the litter from
the jack shafts.
6. With the assistance of another person, lift the litter straight up to remove it from the jack shafts and set it aside.
7. Reverse steps to reinstall the litter.
8. Verify proper operation before returning the unit to service.