www.stryker.com 1089-009-002 REV B 29
Service Information
REMOVAL OF EXCESS AIR FROM THE HYDRAULIC SYSTEM
Tools Required:
• None
Procedure:
1. Verify that all hydraulic linkages are secure and operating properly.
2. Raise the litter to its highest height.
3. Continue to pump the pedal several times to force the air through the system.
4. Verify proper operation before returning the unit to service.