PREVENTATIVE MAINTENANCE PROGRAM
Annual Checklist
____ Inspect for excessive wear the oil-impregnated bronze shoulder spacers found at the bed
hinge points. Replace as needed. Do not lubricate these spacers.
____ Inspect and lubricate when needed the bed lubrication points described in section 2.2 of the
FL17E Maintenance Manual.
____ Inspection of all bolt, locknut and screw tightening. Tighten if necessary.
____ Engage the brake pedal on both sides of the bed (see page 12) and ensure the braking system
operates properly. Toggle the pedal to neutral and ensure the brakes are released.
____ Engage the steer pedal on both sides of the bed (see page 12) and ensure the 5th steer wheel
operates properly. Toggle the pedal to neutral and ensure the steer wheel disengages.
____ Siderails move, latch and stow properly (see page 13).
____ All functions on the foot board control panel (see page 15) working properly (give special
attention to lockout switch LEDs).
____ All functions of the siderail outer/inner control panels (see page 14) working properly. Ensure
the nurse call signal (if present) reaches the nurse station.
____ Optional CPR emergency release handles working properly (see page 18). The Fowler and
Knee Gatch lower completely and the Fowler motor resets itself automatically once the Fowler
is down.
____ Wait about 30 seconds - the time for the Fowler motor to reset itself - then raise the
Fowler to check that the Fowler motor has indeed reset itself.
____ Verify the Fowler and Knee Gatch movements to ensure the motor course is properly adjusted.
Refer to Caution following step 11 of the "Thigh actuator" and the "Head actuator" replacement
procedures found at page 38 and 39 respectively of the FL17E Maintenance manual.
____ Optional 120 volt auxiliary outlet working properly.
____ Optional photoelectric night light working properly.
____ Optional Auto Contour working properly.
____ Foot prop rod working properly when Knee Gatch function is activated.
____ No cracks or splits in boards and siderails.
____ Head end bumpers tightly secured to frame and working properly.
____ No rips or cracks in mattress cover. Remove from service if damaged.
____ On/Off switch and associated LED working properly.
____ Power cord not frayed.
____ No cables worn or pinched.
____ All electrical connections tight.
____ All ground secure to the frame.
____ All casters roll properly. Verify caster tire for cuts, wear.
____ Ground chain intact and in place.
____ Measure current leakage and grounding continuity of the bed and the optional 120V auxiliary
outlet. Verify with our Technical Service department (see section 1.2) for the acceptable
values for this bed.
N
N
O
O
T
T
E
E
We recommend that the bed actuator tubes be replaced after 10 years of service (see
Recommended Spare Parts on page 13 of the Maintenance Manual).
Preventive maintenance may need to be performed more frequently based on the usage level of the
bed.