11.Using a #2 Phillips screwdriver, remove and save the two screws that secure the inspection cover to the base frame
weldment. Remove and save the inspection cover.
12.Using a #2 Phillips screwdriver, remove and save the two screws that secure the board assembly to the base frame
weldment. Remove and save the board assembly.
13.Disconnect the load cell assembly and remove the cable from the base frame weldment. Discard the load cell assembly.
NNoottee -- Note the routing of the load cell cable for reinstallation.
14.Reverse steps to reinstall.
15.Recalibrate the scale (
Calibrate scale
(page 22)).
16.Verify proper operation of the product before you return it to service.
SSiiddeerraaiill rreeppllaacceemmeenntt,, hheeaadd eenndd
TToooollss rreeqquuiirreedd::
T15 Torx driver
T27 Torx driver
Needle nose pliers
Syn-Tech grease (3000-200-719)
PPrroocceedduurree::
1. Raise the product to the highest height position.
2. Raise the Fowler to the full upright position.
3. On the footboard control panel, press EEnntteerr to access the SSeettttiinnggss menu.
4. Highlight SShhuuttddoowwnn and press EEnntteerr.
5. Unplug the power cord from the wall outlet.
6. Push down on the brake pedal to apply the brake.
7. Expand all litter decks to the 48 in. position.
8. Raise the siderail to the up and locked position.
9. Using a T15 Torx driver, remove the two screws that secure the junction board assembly to the siderail carrier. Remove
the junction board assembly.
10.Unplug the two siderail cables from the junction board assembly.
11.Using a T27 Torx driver, remove the screw that secures the guide rod to the litter.
12.Push the siderail release button to lower the siderail to the lowest stowed position.
13.Using needle nose pliers, from under the litter surface, slide each glide rod out.
NNoottee -- Save the two rubber washers on each glide rod, one on both sides of the siderail carrier, to reinstall. Use Syn-
Tech grease to lube the glide rod, if required.
14.While supporting the rail with one hand, remove both glide rods and remove the siderail assembly.
15.Reverse steps to reinstall.
16.Verify proper operation of the product before you return it to service.
SSiiddeerraaiill rreeppllaacceemmeenntt,, ffoooott eenndd
TToooollss rreeqquuiirreedd::
T15 Torx driver
T27 Torx driver
5900-009-002 Rev A.0 33 EN

Other manuals for Stryker MV3 5900

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker MV3 5900 and is the answer not in the manual?

Stryker MV3 5900 Specifications

General IconGeneral
BrandStryker
ModelMV3 5900
CategoryMedical Equipment
LanguageEnglish

Related product manuals