www.stryker.com KK-7100 REV 02 1-17
Operation
Plugging or unplugging the product
WARNING
To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective earth.
Always handle the power cord to avoid the risk of entanglement, damage to the power cord, or potential shock
hazards. If the power cord is damaged, remove the product from service and contact the appropriate maintenance
personnel.
Always allow enough clearance between the head end of the product and the adjacent wall, so you can unplug the
power cord from the wall outlet during an emergency.
Always store the power cord before transporting the product.
To plug in the product, plug the power cord into a protective earthed outlet.
To unplug the product, grasp the mold near the outlet and pull the cord in a direction parallel to the floor (not at an angle).
Charging the battery
WARNING
Always disconnect the power cord from the wall outlet if you detect overheating of the battery, control cables, or
pendants. Do not use the product again until after it has been inspected, serviced, and confirmed to be working to
specification by an authorized and trained maintenance personnel.
Always replace the battery after it surpasses its expected service life.
Do not open a dead battery.
Do not throw the battery into a fire.
Do not spill liquid onto the battery or submerge the battery in liquid.
CAUTION
Always plug the product into a wall outlet (regulated AC-power source) when not in use to maintain a sufficient
battery charge and to maximize product performance while operating on battery power.
Always replace batteries that have corrosion at the terminals, display cracking, have expanded or bulging sides, or
no longer can maintain a full charge.
Only use authorized batteries when replacing the batteries. Use of unauthorized batteries may lead to unpredictable
system performance.
SV1 is equipped with a battery backup system that charges when the product is plugged into a wall outlet. The battery
backup system allows the operator to use the product when the product is unplugged, during a power failure, or while
transporting a patient. Battery backup functionality activates when you unplug the product.
Always check battery backup function according to the preventive maintenance checklist (see Preventive maintenance
on page 1-42). Always replace the battery if there is a failure or malfunction during preventive maintenance.
To charge the battery, connect the product to a wall outlet. The battery has a full charge within 10 to 12 hours.
Storing the battery long term
WARNING
Always unplug the battery cable from the control box before storing the product for a long period of time.

Table of Contents

Other manuals for Stryker SV1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker SV1 and is the answer not in the manual?

Stryker SV1 Specifications

General IconGeneral
BrandStryker
ModelSV1
CategoryMedical Equipment
LanguageEnglish

Related product manuals