EasyManuals Logo

Stryker T2 User Manual

Stryker T2
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #29 background imageLoading...
Page #29 background image
Operative Technique
Fig. 40
Now attach the Paddle Trocar,
Antegrade and the AO T-Handle
Medium Coupling (Fig. 39). Then,
advance them together with the Long
Tissue Protection Sleeve through
the targeting hole for the Antegrade
Femoral Mode (left or right) by
pressing the safety clip (Fig. 40). The
mecha nism will keep the sleeve in
place and prevent it from falling out.
It will also prevent the sleeve from
sliding during screw measurement. To
release the Tissue Protection Sleeve,
the safety clip must be pressed again.
A small skin incision is made and the
assembly is pushed through by mani-
pulating the T-Handle until the Tissue
Protection Sleeve is in contact with the
lateral cortex (Fig. 41).
Guided Locking for Antegrade Femoral Mode
Fig. 41
Fig. 41
Fig. 39
Now attach the Paddle Trocar,
Antegrade and the AO T-Handle
Medium Coupling (Fig. 39). Then,
advance them together with the Long
Tissue Protection Sleeve through
the targeting hole for the Antegrade
Femoral Mode (left or right) by
pressing the safety clip (Fig. 40). The
mecha nism will keep the sleeve in
Guided Locking for Antegrade Femoral Mode
Fig. 39
29