«Definition» Window (commands by dimension with «n» = 1 or 2 (dimension number))
Item Read
command
Write
command
Comment
Resolution nRES? nRES=x x=1 à 5 (number of decimal places)
x=0(mm)
x=1 (inch)
x=2 (µm)
Unit nUNIT? nUNIT=x x=0 (static)
x=1 (maxi)
x=2 (mini)
x=3 (maxi-mini)
x=4 (mean)
x=5 (median)
Dynamic mode nDYN? nDYN?=x x=0 (static)
Formula nFM? nFM=x x=0 (C1)
x=1 (C2)
x=2 (-C1)
x=3 (-C2)
x=4 (C1+C2)
x=5 (C1-C2)
x=6 (-C1+C2)
x=7 (-C1-C2)
The D70 uses the ASCII communications protocol in order to run and configure its functions.
Commands
General details
All commands must terminate with a «CR» character (ASCII code 13). Les commandes
peuvent être envoyées groupées en les séparants par un caractère «;» (max 500 caractères).
L’affichage est rafraîchi une seule fois après la ception du caractère «C.
List of commands
DEFINITION window
Communication protocol
17