Chapter 2 Safety Information
SA-90 Instructions for Use
14/
46
2.2 Installation
2.3 Electromagnetic compatibility (EMC)
This instrument complies with the following IEC (EN) standards:
• IEC61326-2-6:2012 (EN61326-2-6:2013)
Electrical equipment for measurement, control and laboratory use - EMC requirements
• EMI (Electromagnetic Interference) For this standard the requirements of class A are fulfilled.
• EMS (Electromagnetic Susceptibility)
For this standard the minimum requirements with regards to susceptibility are fulfilled.
• This equipment has been designed and tested to CISPR11 Class A. In a domestic environment it may cause a
radio interference, in which case, you may need to take measures to mitigate the interference. The electromagnetic
environment should be evaluated prior to operation of the device. Do not use this device in close proximity to
sources of strong electromagnetic radiation (e.g. unshielded intentional RF sources), as these may interfere with
the proper operation.
• The unpacking of the instrument, installation, and initial operation checks will be performed by your
authorized local representative.
• This instrument must not be connected to a power outlet rated at anything other than AC 100 to 240 V.
The instrument must be grounded.
Otherwise, fire or electric shock may result.
• When connecting accessories and peripherals, always turn the power off first.
Otherwise, electrical shock or instrument failure may result.
• Install the instrument in a location where it will not be exposed to water.
• Install the instrument in a location where it will not be adversely affected by heat, humidity, dust, or direct
sunlight.
• Make sure the instrument is not exposed to strong vibrations or impacts.
• Install the instrument in a well-ventilated area.
• Do not install the instrument near devices that emit electrical interference, such as a radio or centrifugal
device.
• Do not install the instrument close to stores of chemical substances or other sources of gas emissions.
• Do not use this instrument in any operating environment which has electroconductive gases, flammable
gases, oxygen, hydrogen, or anesthetics that include other flammable gases.
• Install the instrument in an indoor environment.
This instrument is designed for indoor use only.
• Once the installation is completed, securely fix the instrument by locking adjuster feet.