EasyManua.ls Logo

SYSMEX SA-90 - Chapter 7 Basic Operation; Flow of Basic Operation

SYSMEX SA-90
46 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 7 Basic Operation
SA-90 Instructions for Use
31/
46
Chapter 7 Basic Operation
7.1 Flow of basic operation
The following flowchart shows the outline of operations for the instrument. The typical flow is shown in gray.
Note:
To use the instrument, connection settings must be configured on the CN-3000, CN-6000, CN-3500 or
CN-6500.
(³CN-3000 and CN-6000 Instructions for Use, "Chapter 13: 13.2.22 Option sampler connection settings")
(³CN-3500 and CN-6500 Instructions for Use, "Chapter 13: 13.2.22 Option sampler connection settings")
Start the CN-3000, CN-6000, CN-3500 or CN-6500
(³CN-3000 and CN-6000 Instructions for Use, "Chapter 7: 7.2 Starting the instrument")
(³CN-3500 and CN-6500 Instructions for Use, "Chapter 7: 7.2 Starting the instrument")
7.2 Startup ³P. 32
7.2.3 Executing mechanical initialization (approx. 2 minutes) ³P. 32
Perform sample analysis
(³CN-3000 and CN-6000 Instructions for Use, "Chapter 7: Basic Operation")
(³CN-3500 and CN-6500 Instructions for Use, "Chapter 7: Basic Operation")
7.4.1 Placing a sample rack in the rack feeder ³P.3 3
7.4.2 Removing the sample rack once it has been transported to the rack collector ³P. 34