M503 Series chair Scales Service Manual (3.09)
- 15 -
Cap2: capacity of second range
cal
Enter calibration
Dual RA
desc
Set decimal point
Inc
Set increment of dual range
Sdiv 1increment of first range
Sdiv 2
increment of second range
cap
Set capacity of dual range
Cap1: capacity of first range
Cap2
: capacity of second range
cal
Enter calibration
P 3 pro
tri
Display will show XXXXX. For trimming the load
cells, showing primary weight.
You can calculate new rate by this formula:
N2=N1+N1×[(K2-K1)÷K2]
N1: primary rate, N2: new rate, K1: calibrate weight,
K2: display weight
count
To show the scale internal count
reset
Reset the scale
setgra
Set the gravity value
8. CALIBRATION
Calibration Settings in the Parameter;
Turn on the scale. And when in the normal display
Press and hold
key, display will show
Press key , display will show
Press key to confirm, display will show
Press and keys orderly, display will show
Press key, display will show
F1 off
tch
pin
P1 spd
P2 cal