EasyManua.ls Logo

Tanaka MPC-102L - Setting the Coolant Temperature (MPC-102 L Only); Specimen Preparation

Tanaka MPC-102L
36 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
- 15 -
5.2. Setting the Coolant Temperature (MPC-102L Only)
!
Caution: Never press the START key unless coolant is circulating
normally. Starting the unit with no coolant present may
damage the thermo-module that cools and heats the
bath.
Start the coolant circulation device and enter a temperature (refer to the Coolant Circulation Device
Instruction Manual). Since the lower limit of the measuring range of the instrument depends on the
coolant temperature (see Table 2), ensure that you have selected the correct temperature for the
specimen you are testing.
Table 2 Coolant Circulation Device Set Temperature
Coolant circulation device set
temperature
Measuring range
0 51 to -40
-40 51 to -65
5.3. Specimen Preparation
Fig. 7 Cross Section of the Detector Assembly
Pressure conductin
g
tube
Tem
p
erature senso
r
Cloud point sensor
(
Li
g
ht
g
uide
)
Compound
Detecto
r
Metallic ti
p
O Rin
g

Related product manuals