- 32 -
pour/cloud point. At shipping, the switching temperature is set to 40°C above the
Expected pour/cloud point.
Safety and self-diagnosis systems:
1. If the temperature on the hot side of the Peltier cooler exceeds 60°C,
a warning buzzer sounds and the power supply is automatically cut to
prevent fire hazards.
2. The following error conditions are automatically detected, and
displayed on the main screen, accompanied by an intermittent buzzer:
・ Temperature sensor line cuts are detected.
・ The intensity of light detected by the cloud point detection sensor is
incorrect.
・ The internal battery that maintains internal memory settings is
getting flat. (Tested at power-on only)
3. Computer runaway (Buzzer sounding only)
・ Some kind of circuit failure is detected.
Data Outputs: Supporting LAS RS-232C × 1ch (standard)
Remarks: One additional channel available as an option (program
alteration necessary).
Power Supply: 100/120 or 220/240 VAC (Factory setted) 50/60 Hz 0.5 kW
Dimensions and Weights: Control unit: W230×D455×H110 mm Approx. 6.0 kg
Test unit: W230×D480×H275 mm Approx. 18.0 kg
Minimum Footprint: W300×D550 mm
Standard Accessories:
Specimen Cup with Reflex Seal ································································································· 5 pcs
Insulator Pad ······························································································································ 1 pc
Spare Pressure Conducting Tube ······························································································ 1 pc
AC Power Cord ························································································································· 1 pc
Connection cable (power) 0.6m ································································································· 1 pc
Connection cable (signal) 0.6m··································································································1 pc
Instruction manual ······················································································································ 1 copy
Maintenance manual ·················································································································· 1 copy
<Above common to MPC-102A and MPC-102L>
Coolant circulation hose ∅10×∅14 2m [with insulation pipe] ···················································· 2 pcs
Hose Clamps ······························································································································ 4 pcs
Dripping plate ····························································································································· 1 pc
<Above for MPC-102L only>