6.4.4. To Release From Slave Mode
Slave units can be released using the following procedure:
a) Depress FINE button for 3 seconds. The current display changes to show the
master or slave setting.
b) Select “H1” mode using the CURRENT encoder.
c) Depress FINE button again or wait 5 seconds.
d) Turn the AC power Off and On to store the new setting.
e) The front panel will be locked. Unlock it by depressing the PREView button until the
display shows “UFP”.
f) After exiting from Slave operation the unit’s parameters will be set to:
Programmed Voltage to zero AST OFF
Programmed Current to zero OUT OFF
UVL to zero volts Foldback protection OFF
OVP to its maximum value Locked Front Panel