3 - 6 MIO2 Operating Manual, 394934, en, V3.0
3 - Technical Data
Operating Range
3.4 Operating Range
6 Slot MIO
4 Slot MIO
Tab. 3-6 Operating ranges, 6 positions
Number of Slots 6
Incubating tempera-
ture range
from (actual room temperature + 5°C) to 60°C
from (actual room temperature + 9°F) to 140°F
Accuracy of chamber
temperature
max. ±0.5°C
at 37°C (max. ±0.9°F at 99°F)
max. ±0.5°C
at 46°C (max. ±0.9°F at 115°F)
max. ±2.0°C at 60°C (max. ±3.6°F at 140°F)
Accuracy of tempera-
ture in microplate in
the chamber
max. ±1°C at 37°C (max. ±1.8°F at 99°F)
max. ±1°C at 46°C (max. ±1.8°F at 115°F)
max. ±2°C at 60°C (max. ±3.6°F at 140°F)
Warm-up time from
20°C to 37°C
max. 20 min.
Warm-up time from
20°C to 45°C
max. 20 min.
Warm-up time from
20°C to 60°C
max. 30 min.
Shaking / rotation fre-
quency
1–8.5 Hz
Linear shaking max. 2 mm amplitude in Y-dimension
Tab. 3-7 Operating ranges, 4 positions
Number of Slots 4
Incubating tempera-
ture range
from (actual room temperature + 5°C) to 60°C
from (actual room temperature + 9°F) to 140°F
Accuracy of chamber
temperature
max. ±1°C at 37°C (max. ±1.8°F at 99°F)
max. ±1°C at 46°C (max. ±1.8°F at 115°F)
max. ±2°C at 60°C (max. ±3.6°F at 140°F)
Accuracy of tempera-
ture in microplate in
the chamber
max. ±1°C at 37°C (max. ±1.8°F at 99°F)
max. ±1°C at 46°C (max. ±1.8°F at 115°F)
max. ±2°C at 60°C (max. ±3.6°F at 140°F)
Warm-up time from
20°C to 37°C
max. 30 min.