7 - 2 MIO2 Operating Manual, 394934, en, V3.0
7 - Maintenance
General
7.3 General
WARNING
Injuries or contamination with hazardous materials possible if the maintenance is
not carried out properly.
Make sure that the instrument worktable and the MIO2 are free from
hazardous materials. Decontaminate them (or have them decontaminated) if
necessary.
Protect your body with adequate protective clothing, goggles and gloves.
Make sure MIO2 has cooled down to room temperature before performing
maintenance tasks.
Make sure the incubator slots are empty before performing maintenance
tasks.
Once installed, the MIO2 is part of the instrument.
Decontaminate and clean the MIO2 when decontaminating and cleaning the
instrument.
For detailed information on maintenance of the instrument, refer to the Operating
and Service Manuals of the instrument. Refer to section 1.7 “Reference
Documents”, 1-2.
7.4 Special Maintenance Tasks
7.4.1 Incubator Removal
1 Make sure the electrical power of the instrument is switched off.
ATTENTION
Risk of damage to equipment. Incubators that are installed in disorder will lead to
damage of equipment.
Only remove one incubator at the time if multiple incubators are installed to
avoid muddling up.
2 Lift and turn the incubator to remove the D-Sub cable connected to the MIO2.
3 Completely remove the incubator.