and the symbol bar - will be displayed on the LCD.
The lowest detect voltage is around 50V 50/60Hz.
The LCD display EF, when detect the AC voltage signal, the
LCD display - ’/‘ -- ’/‘ --- ’/‘ ---- from weak to strong.
This function is depending model.
3.2.10 PC Link
The meter has serial data output function. It can be
connected with PC by USB interface, so the measured data
can be recorded, analyzed, processed and printed by PC.
Before use this function, you need install the PC-Link
software and USB driver in your PC.
Keep press REL more than 2 seconds and the meter enter
PC-Link mode, the symbol “PC-LINK” will appear on LCD,
and the serial data output function is active.
Auto power off function is disable when PCLINK function
active.
PC-LINK SOFT OPERATING MANUAL
1. Make sure the two Install USB driver and Install
software/RUN files in the attached CD successfully
installed before any measurement.
2. Connect the meter’s OPTICAL PORT and computer USB
port with the USB cable.
3. Run the PC-LINK software, click the Start button.
4. Keep press REL key more than 2 seconds on the meter,
the symbol PC-LINK will appear on the LCD if the serial
data output function is active.
As for check the USB cable success recognized, we can