EasyManua.ls Logo

Tektronix 11801C - Table B 6: 11801 C AC Line Power Specifications; Table B 7: 11801 C Environmental Specifications

Tektronix 11801C
440 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix B: Specifications
Appendices
BĆ6
TableĂBĆ6:ă11801C AC Line Power Specifications
Characteristic Specification
Voltage ranges 90 to 132 V
RMS
or 180 to 250 V
RMS
Voltage ranges apply to trace distortion,
which reduces peak line voltage 5%
Installation Category II
Frequency 48 Hz to 440 Hz
Power 320 W
Current 4.6 A
RMS
at 50 Hz, 90 V line with 5%
clipping
Fuse 6 A, 250 V normal blow
TableĂBĆ7:ă11801C Environmental Specifications
Characteristic Specification
Temperature Meets MIL-T-28800E, Type III, Class 5
Operating 0_ C to 50_ C
NonĆoperating -40_ C to 75_ C
Humidity Up to 95% relative humidity, at up to 50_ C
Altitude Meets MIL-T-28800E, Type III, Class 5
Operating Up to 4.5 km (15,000 ft)
NonĆoperating Up to 15 km (50,000 ft)
Vibration Operating, sampling heads not installed:
meets MIL-T-28800E, Type III, Class 5 (secĆ
tion 4.5.5.3.1)
Shock NonĆoperating, sampling heads not installed:
meets MIL-T-28800E, Type III, Class 5
(section 4.5.5.4.1)
Bench handling Operating: meets MIL-T-28800E, Type III,
Class 5 (section 4.5.5.4.3)
Packaged product vibration
and bounce
Packaged product, sampling heads not
installed: meets NSTA Proj. 1A
Drop of packaged product Packaged product, sampling heads not
installed: meets NSTA Proj. 1A
Electrostatic immunity No disruption or degradation of performance:
up to 10 kV. No damage to instrument: up to
20 kV

Table of Contents