EasyManuals Logo
Home>Tektronix>Test Equipment>2215A

Tektronix 2215A Operator's Instruction Manual

Tektronix 2215A
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #31 background imageLoading...
Page #31 background image
t
NEGATIVE
REFERENCE
LINE
-----------
--
--
J
VERTICAL
DEFLECTION
............... .
-----
,----
POSITIVE
REFERENCE
LINE
MEASURE
POSITIVE
AMP
LI
TUDE
TO
OR
NEGATIVE
AMPLI
T
UDE
TO
4732·08
Figure
4-2
. Instantaneous vol
tage
meas
urement.
10. Calculate the instantaneous voltage, using the fol-
lowing formula:
lnstanta-
neous
Voltage
vertical
deflection x polarity x
(divisions) (
+
or
- )
VOLTS/DIV
switch setting
indicated by 1 X
(or 1
OX
when 1
OX
probe is used)
EXAMPLE: The measured vertical deflection from the
reference line is 4.6 divisions (see Figure 4-2), the wave-
form point is above the reference line, a 1
OX
attenuator
probe is be
in
g used, and
th
e VOLTS/DIV switch is set
to
2 V (at 1
OX
PROBE setting).
Substituting the given values:
I
nstantaneous
Voltage
=
4.6
div x ( +
1)
x 2 V/
div
=
9.2
V.
Algebraic Addition
With the VERTICAL MODE switches set
to
BOTH and
ADD, the waveform displayed is the algebraic sum
of
the
signals applied
to
the Channel 1 and Channel 2 inputs (CH 1
+
CH
2).
If
the Channel 2 INVERT button is pressed in, the
waveform displayed is the difference between the signals
applied
to
the Channel 1 and Channel 2 inputs (CH 1 -
CH
2).
When both VOLTS/DIV switches are set
to
the same
deflection factor, the deflection factor in the ADD mode is
equal
to
the deflection factor indicated by either VOLTS/DIV
switch.
The following general precautions should be observed
when using the ADD mode.
Operating Procedures
-22
15A Operators
a.
Do
not
exceed the input voltage rating
of
the
oscilloscope.
b.
Do not apply signals that exceed the equivalent of
about eight times the VOLTS/DIV switch settings,
since large voltages may distort the display. For ex-
ample, with a VOLTS/DIV switch setti
ng
of
0.5 V, the
voltage app
li
ed
to
that channel should not exceed ap-
proximately 4 volts.
c.
Use Channel 1 and Channel 2 POSITION control
settings which most nearly position the signal on each
channel
to
midscreen, when viewed in either
CH
1
or
CH 2 VERTICAL MODE. This ensures the greatest
dynamic range for ADD mode operation.
d.
To
attain similar response from each channel, set
both the Channel 1 and Channel 2 Input Coupli
ng
switches
to
the same position.
Common-Mode Rejection
The following procedure shows how
to
eliminate
un
-
wanted ac input-power frequency components. Similar
methods could be used either
to
eliminate other unwanted
frequency components
or
to provide a de offset.
1 . Preset instrument controls and obtain a baseline
trace.
2. Apply the signal containing the unwanted line-
frequency components
to
the
CH
1 input connector.
3. Apply a line-frequency signal
to
the CH 2 input con-
nector.
To
maximize cancellation, the signal applied
to
Channel 2 must be in phase with the unwanted line-
frequency component on the Channel 1 input.
4. Select BOTH and ALT VERTICAL MODE
and
set
both VOLTS/DIV switches
to
produce displays
of
approxi-
mately 4-
or
5-divisions in amplitude.
5. Adjust either VOLTS/DIV switch
and
VOLTS/DIV
Variable control so that both display signals are the same
amplitude.
6. Adjust the
CH
2 VOLTS/DIV switch and
CH
2
VOLTS/DIV Variable control so that the Channel 2 display is
approximately the same amplitude as the undesired portion
of
the Channel 1 display (see Figure 4-3A).
4-3

Table of Contents

Other manuals for Tektronix 2215A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2215A and is the answer not in the manual?

Tektronix 2215A Specifications

General IconGeneral
BrandTektronix
Model2215A
CategoryTest Equipment
LanguageEnglish

Related product manuals