EasyManuals Logo

Tektronix 2215A Operator's Instruction Manual

Tektronix 2215A
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #39 background imageLoading...
Page #39 background image
I
NTENS
I
FIED
/
ZONE
A
TRACE
B
TRACE
r
-
ALT
HORIZ
DISPLAY
\
VERTICAL
REFERENCE
LINE
C4207-23
Figure
4-12.
Time
difference
on
single
waveform.
9. Adjust the B DELAY POSITION control clockwise
to
move the intensified zone
to
the leading edge
of
the second
point
of
interest (on the A trace); then fine adjust until the
rising portion
(on
the B trace) is centered at the same conve-
nient vertical graticule used in preceding step
7.
10. Record the B DELAY TIME POSITION control dial
setting.
11
. Calculate the time difference between repetitive
pulses using the following formula.
Time (second first )
(A
SEC/
DIV)
Difference = dial - dial switch
(Duration) setting setting setting
EXAMPLE: With the A SEC/DIV switch set
to
0.02 ms
per divisions the first B DELAY TIME POSITION dial set-
ting is 1.20 and the second B DELAY TIME POSITION
dial setting is 9.53 (see Figure 4-13).
Substituting the given values
in
the time difference
formula:
Time Difference = (9.53 - 1.20) (0.2 ms/div) = 1.666 ms
Operating Proce
dures
-22
15A
Operators
DIAL
READING
AT
PO
IN
T A
DIAL
READIN
G
AT
POINT
B
C4207-24
Figure 4-13. B DELAY TIME POSITION
control
setti
ngs
.
Rise Time
Rise-time measurements use the same methods as time
interval on single waveform, except that the measurements
are made between the 1 0% and 90% points
of
the low
to
high transition
of of
the selected waveform. Fall ti
me
is mea-
sured between the 90% and 10% points
of
the high
to
low
transition of the waveform.
1 . Preset instrument controls and obtain a baseline
trace.
2. Apply an exact 5-division signal
to
either vertical-
channel input connector and set the VERTICAL MODE
switch
to
display the channel used.
3. Set the appropriate VOLTS/DIV switch and variable
control for an exact 5-division display.
4. Vertically position the trace so that the zero reference
of
the waveform touches the 0% graticule line and the top
of
the waveform touches the 100% graticule line.
5.
Set the A SEC/DIV switch
for
a single-waveform dis-
play. Ensure that the
SEC
/DIV Variable control is
in
the CAL
detent.
6. Select
ALT
HORIZONTAL MODE and set the
B
SEC
/DIV switch
to
spread the rise-time-measurement
portion
of
the display as much as possible.
4-11

Table of Contents

Other manuals for Tektronix 2215A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2215A and is the answer not in the manual?

Tektronix 2215A Specifications

General IconGeneral
BrandTektronix
Model2215A
CategoryTest Equipment
LanguageEnglish

Related product manuals