Controls,
 Connectors,
 and
 indicators 
Activates the Delta Time measurement function and 
cancels the AV or 1/At measurement functions. When At 
is selected with Inten or Alt horizontal display modes, 
two delay times are
 defined.
 When At is selected with 
either A-Sweep or
 B-Sweep
 horizontal display, two 
vertical cursors are established. One delay time or 
cursor position is controlled by the A REF OR DLY POS 
control,
 and the other is controlled by the A control. The 
CRT readout displays either the difference between the 
two delay times or the equivalent time between the 
vertical cursors. 
If SEC/DIV VAR is not in the detent position, and either 
the A-Sweep or the
 B-Sweep
 horizontal display mode is 
selected,
 the CRT readout displays delta-time as a ratio, 
where five divisions correspond to 100% ratio. 
When At is active, pressing the At button deactivates 
the function. 
1/At Momentarily pressing the At and AV buttons 
Function together activates the
 1/Delta-Time
 function and 
cancels any other Delta measurement function. The 
waveform display and the Delta controls operate the 
same as for At, but the readout shows the reciprocal 
of the time in Hz (frequency). 
If the SEC/DIV VAR control is not in the detent 
position (full clockwise rotation), and the A-Sweep or 
B-Sweep
 horizontal display mode is selected, the 
readout displays the time between cursors as 
degrees of phase, where five divisions are equal to 
360 degrees. 
When the
 1/Delta
 Time function is active, pressing 
the At and AV buttons together deactivates the 
function. 
34) At Button 
5-16 
2465B/2455B/2445B Operators