Checks and Adjustments 
To reduce channel-to-channel signal delay errors with Parametric 
Measurements: 
1.
 Follow Steps 1-3, above. 
2.
 Measure the time interval from
 CH1
 to CH2 using the TIME function 
from the MEASURE menu. 
3. Use this number as a reference and subtract its value from all 
subsequent time interval measurements. 
Amplitude
 Check 
1.
 Obtain a display as described in "Initial Setup.* 
2.
 Set the VOLTS/DIV switch to 100 mV, the SEC/DIV switch to
 1
 ms, and 
20 MHz BW LIMIT on. 
3. Adjust the CH 1 POSITION control to center the display on the screen. 
4.
 CHECK—Amplitude of the CALIBRATOR signal is between 3.88 and 
4.12 divisions as measured on the center vertical graticule line. 
5. Select AV and carefully superimpose the cursors on the high and low levels 
of the waveform. CHECK—AV readout is between 392 mV and 408 mV. 
6. Repeat this procedure using the Channel 2 connector and controls. 
4-6 
2465B/2455B/2445B Operators