Controls,
 Connectors,
 and
 Indicators 
(62) dBV/dBm Selects or cancels dBV or dBm. 
dBV 
Displays ac rms voltage measurements as the log of 
the ratio of the input voltage to
 1
 V. Readings equal 
20xlog(V
jnput
/V
ref
), where V
jnput
 is the input voltage 
and V
ref
 is
 1
 V. With zero input, the readout 
displays "OVER." 
dBm 
63) CONT 
Displays ac rms voltage measurements as the log of 
the ratio of the input voltage to a
 1
 mW reference at 
600 ft. Readings equal 20xlog(V
jnpu1
/V
ref
), where 
v
m
P
ut
is the
 in
P
ut
 voltage and V
ref
 is
 0.7746
 V. With 
zero input, the readout displays "OVER." 
Selects or deselects continuity. If the resistance between 
the probes is less than 10 ft, the instrument produces an 
audible tone. The readout shows either "SHORT < 
10 ft" or "OPEN > 10 ft." The tone responds to 
continuity within about 100 ms. 
Continuity operation is not affected by AUTO, UP, 
DOWN,
 HOLD, SMOOTH, MIN/MAX DISPLAY, 
MIN/MAX RESET, REF DISPLAY, or REF SET. 
(64) HI ft/LO ft Selects or deselects resistance measurement. 
HI ft 
Measures resistance within 2 k, 20 k, 200 k, 2 M, 
and 20 Mft ranges. At full scale, 2 V appears across 
the test leads, which will bias a silicon semiconductor 
junction to forward conduction. 
LOft 
65) °C/°F 
Measures resistance within 200, 2 k, 20 k, 200 k, 
and 2 Mft ranges. At full scale, 0.2 V appears across 
the test leads, which will not bias a silicon 
semiconductor junction to forward conduction. 
Selects or deselects temperature measurement. If a 
temperature probe is not connected to the input jacks, 
the readout shows "ATTACH TEMP PROBE/' 
Measures temperature in Celsius degrees. 
Measures temperature in Fahrenheit degrees. 
5-30 
2465B/2455B/2445B Operators