Appendix
 C 
Table C-1 
Confidence Test Numbers and Affected Functions 
Test Number 
01 
02 
03 
04 
05 
06 
GP 11 
DM 76 
CT
 81
 - 87 
Description 
Controller Timing Signal Missing or Wrong Period 
Momentary Switch Stuck 
Readout Interface or Memory Failure 
Calibration Data Parity or Checksum Error 
Main Board Failure Detected by Auto Level Trigger on LINE 
Source 
Memory-Battery Voltage Too High or Too Low 
GPIB Interface Failure Detected 
Digital Multimeter Failure Detected 
Counter-Timer-Trigger Failure Detected 
C-2 
2465B/2455B/2445B Operators