EasyManuals Logo

Tektronix 2465B Service Manual

Tektronix 2465B
536 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #166 background imageLoading...
Page #166 background image
Performance Check—2465B/2467B Service
1.
Function Generator (60 Hz, 30 kHz and 80 kHz)
2.
Primary Leveled Sine-Wave Generator (50 MHz)
3. Secondary Leveled Sine-Wave Generator
(500 MHz)
NOTE
To obtain signal amplitudes less than 1
division, first set the signal for either 4, 5,
or 10 times the specified amplitude, then
reduce the amplitude by a factor of
4,
5, or
10 by increasing the VOLTS/DIV settings
as necessary.
c. For each combination listed in the table, set the
gen-
erator Test Frequency and the oscilloscope TRIGGER
COUPLING as indicated, performing the following steps to
verify the Triggering levels in each setup.
d.
Set the VOLTS/DIV and the generator output level to
obtain the test signal amplitude indicated for the particular
combination being tested. When checking channel 1 and
channel 2 500 MHz triggering, also adjust the VOLTS/DIV
VAR for the correct input level.
e. Set the A SEC/DIV and the X10 MAG to obtain a
well-defined display of the test signal.
NOTE
Normally, unless trigger sensitivity is very close to
the specified limits, it is sufficient to check each of
the indicated frequency-coupling combinations listed
in the table in Channel 1
only;
checks for Channels
2, 3 and 4 need only be done in DC COUPLING (to
verify signal path).
f. CHECK—For a stable triggered display (unless other-
wise indicated) for each of the Test Frequency-TRIGGER
COUPLING combinations listed in Table 4-5. When testing
the 300 MHz triggering, check that trigger jitter is
< 100 ps (0.2 division at 5 ns/div with X10 MAG), with 5
divisions of signal and TRIGGER LEVEL adjusted for
minimum jitter.
g.
Press the ADD button to select the function and
press the CH 1 button to turn off the CH 1 display.
h. Repeat the DC TRIGGER COUPLING tests of Table
Table 4-5
CH 1 or CH 2 Triggering Conditions
Test
Fre-
quency
60 Hz
30 kHz
80 kHz
50 MHz
300 MHz
500 MHz
Minimum Vertical Display Levels at Which
Triggering Should Occur
TRIGGER COUPLING
DC
a
a
a
0.35
Div
1.0
Div
1.5
Div
NOISE
REJ
a
a
a
1.2
Div
3.0
Div
4.5
Div
HF REJ
a
0.35
Div
a
No Trigger,
Freeruns
at 1.2 Div
No Trigger,
Freeruns
at 3.0 Div
a
LFREF
No
Trigger,
Freeruns
a
0.35
Div
0.35
Div
1.0
Div
1.5
Div
AC
0.35
Div
a
a
0.35
Div
1.0
Div
1.5
Div
"Not necessary to check.
4-5 while in the ADD mode, adding 0.5 DIV to the 300 and
500 MHz amplitudes.
i. Move the signal to the CH 2 input connector and
repeat step h for CH 2.
j.
Press the CH 2 button to select the channel and
press the ADD button to turn off the ADD display.
k. Repeat the DC TRIGGER COUPLING tests of Table
4-5 while in CH 2 mode.
I. If trigger sensitivity is close to the specified limits
given in steps c through k above, test all of the
frequency-coupling combinations given in Table 4-5 for
CH 2.
m. Move the test signal to CH 3 and CH 4 in turn and
repeat parts c through f using Table 4-6.
4-16

Table of Contents

Other manuals for Tektronix 2465B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2465B and is the answer not in the manual?

Tektronix 2465B Specifications

General IconGeneral
Bandwidth400 MHz
Channels4
Vertical Sensitivity2 mV/div to 5 V/div
Maximum Input Voltage400 V (DC + peak AC)
DisplayCRT
Trigger ModesAuto, Normal, Single
Time Base Range500 ps/div to 0.5 s/div
Rise Time875 ps
Power Supply48 to 440 Hz
Dimensions495 mm

Related product manuals