Performance Check—2465B/2467B Service 
Table 4-8 
Horizontal Timing Accuracy Checked Against the Graticule 
Time-marker Accuracy 
(X10 MAG off) 
Time-marker Accuracy 
(X10 MAG on) 
(Exclude first 0.5 division 
of sweep rate) 
As Measured Against These Time-
Marker Pairs (X10 MAG off only) 
Over Any 
1 Div 
±0.07 
Div 
±0.07 
Div 
1-2 
2-3 
3-4 
4-5 
5-6 
6-7 
7-8 
8-9 
9-10 
10-11 
2 Div 
±0.07 
Div 
±0.08 
Div 
1-3 
2-4 
3-5 
4-6 
5-7 
6-8 
7-9 
8-10 
9-11 
3 Div 
±0.08 
Div 
±0.1 
Div 
1-4 
2-5 
3-6 
4-7 
5-8 
6-9 
7-10 
8-11 
4 Div 
±0.09 
Div 
±0.11 
Div 
1-5 
2-6 
3-7 
4-8 
5-9 
6-10 
7-11 
5 Div 
±0.10 
Div 
±0.12 
Div 
1-6 
2-7 
3-8 
4-9 
5-10 
6-11 
6 Div 
±0.10 
Div 
±0.13 
Div 
1-7 
2-8 
3-9 
4-10 
5-11 
7 Div 
±0.11 
Div 
±0.14 
Div 
1-8 
2-9 
3-10 
4-11 
8 Div 
±0.12 
Div 
±0.16 
Div 
1-9 
2-10 
3-11 
9 Div 
±0.12 
Div 
±0.17 
Div 
1-10 
2-11 
10 Div 
±0.13 
Div 
±0.18 
Div 
1-11 
g.
 Rotate the A REF OR DLY POS control CCW until 
the cursor stops moving. 
h. CHECK—A REF OR DLY POS cursor aligns with the 
1st graticule line within 0.2 division. 
i. Rotate the A control CW until the cursor stops 
moving. 
j.
 CHECK—A cursor aligns with the 11th graticule line 
within 0.2 division. 
k. Set the A SEC/DIV to 10 ns. 
I. Rotate the A REF OR DLY POS and the A controls to 
precisely superimpose the cursors near the 2nd graticule 
line. 
m. CHECK—At readout indicates a difference of 
0.30 ns or less. 
4-23 
n. Rotate the A REF OR DLY POS and the A controls 
to precisely superimpose the cursors near the 10th grati-
cule line. 
o. CHECK—At readout indicates a difference of 0.30 ns 
or less. 
p. Set: 
B SEC/DIV 5 ns (knob in) 
B TRIGGER MODE RUN AFT DLY 
X10MAG Off 
At Off (DLY) 
A REF OR DLY POS Set for zero delay 
NOTE 
On CTT instruments, rotate the A REF OR DLY 
POS control for the specified delay. As the 
control is rotated, the readout delay value will be 
followed by the word
 "SET".
 This value shows 
the approximate delay. A few seconds after con-
trol movement has stopped, the word "SET" will 
disappear and the readout delay value as meas-
ured by the CTT will appear. This is normal 
operation and not cause for concern.