Performance Checkā2465B/2467B Service 
q.
 CHECKāThe B sweep timing accuracy as in parts b 
through f, making sure that the A SEC/DIV is set slower 
than the B SEC/DIV. 
2.
 Select COUNTER ("4") from menu. 
3. Select PERIOD ("2") from menu. 
3. Check Delta Time Accuracy using the Delayed 
Sweep. 
a. Set: 
A SEC/DIV 
B SEC/DIV 
X10 MAG 
At 
TRIGGER MODE 
TRIGGER SOURCE 
TRIGGER COUPLING 
TRIGGER SLOPE 
TRIGGER LEVEL 
BTRIG MODE 
10 ns 
5 ns (knob out) 
On 
Off (DLY readout) 
AUTO LVL 
VERT 
DC 
+ (plus) 
As required for a 
stable display 
RUN AFT DLY 
NOTE 
Certain time marks from the TG 501 (and other 
Time-Mark Generators) will vary in width and may 
be displaced in time. This will happen in a repeat-
able sequence and is caused by the loading and 
interaction of the 2, 5, and 10 dividers. This is 
most noticeable with 10 ns, 20 ns, and 50 ns 
markers. The following procedure will use the 
above markers to set up the proper references 
but the 5 ns markers will be used to make the 
actual measurement. Close inspection of apparent 
jitter or mistrigger of the time marks will show the 
trigger point to be stable with the apparent jitter 
to be variable with unique combinations of trigger 
holdoff and sweep speed. This is normal behavior 
with this type of signal and is not an instrument 
defect. 
The CTT period readout will appear on the left side of 
the upper line of readout. The word "SET" will appear next 
to the readout delay value. This denotes the indirect mea-
surement mode of Delta Time, simulating a non-CTT 
scope. 
b. Set the Time-Mark Generator for 10 ns markers. 
Adjust the Vertical VOLTS/DIV as required for a display of 
3 to 6 divisions. 
c. Adjust the A REF OR DLY POS control for a readout 
display of DLY 10.64 ns. 
d.
 Adjust the Horizontal POSITION control CW until the 
trace stops moving, then CCW to display the leading edge 
of the 2nd time marker near the graticule center. This 
becomes the reference point for the following procedure. 
Set the Time-Mark Generator to 5 ns and adjust the Verti-
cal VOLTS/DIV and Trigger LEVEL as required. 
It is not necessary to count the number of marks given 
in the tables. Switching to 10 ns, 20 ns, or 50 ns markers 
as required and then to 5 ns will show the proper 5 ns 
mark to be used. 
e. Press and release the At button to obtain the At 
display. Push in the SEC/DIV knob for B SWP only. 
Rotate the A control for a readout display of At 
-10.64 ns. If the time marks are not superimposed, adjust 
the A control to do so. 
For CTT instruments, use the following setup of the 
CTT while performing the Delta Time performance check. 
1.
 Push the MEASURE button to enter MENU 
mode. 
f. CHECKāAt readout is within the limits listed in Table 
4-9 for the 1st 5 ns time marker; then check that the 3rd 
through 19th time markers are within the given limits as 
the A control is rotated CW to superimpose every second 
time marker on the reference time marker. 
4-24