Commands Listed in Alphabetical Order
Examples
COMPENSATE:DA
TE:CH2?
might return COMPENSATE:D ATE:CH2 15 JAN 00
16:25
.
COMPensate:DATE:MAInframe? (Query Only)
This is a query only command that returns the date and the time of the current
in-use (that is, run-time) compensation data for the mainframe.
Group
Compensation
Syntax
COMPensate:DATE:MAInframe?
Examples
COMPENSATE:DATE:MAINFRAME? might return
COMPENSATE:DATE:MAINFRAME 15 JAN 00 16:25.
COMPensate:MAInframe (No Query Form)
This command (no query form) compensates the mainframe for DC variances.
Before sending this command, disconnect or terminate all sampling module
inputs, and disconnect all probes.
NOTE. Data from a successful compensation is not automatically saved into
nonvolatile user storage. Use the COMPensate:SAVe commands if you want to
retain the data in nonvolatile storage.
Group
Compensation
Syntax
COMPensate:MAInframe
Related Commands
COMPensate:SAVe:USER:MAInframe
Examples
COMPENSATE:MAINFRAME performs the compensation routine on the mainframe.
COMPensate:OPTGAIN:CH<x> (No Query Form)
This command (no query form) compensates the gain of the optical channel by
comparing the given input power to an average optical power measurement. The
gain correction factor and the given input wavelength are automatically stored
DSA/CSA/TDS8X00/B Series Programmer Manual 2-93