Commands Listed in Alphabetical Order
ACQuire? (Query Only)
This is a quer
y only command that returns all the current acquisition parameters.
Group
Acquisition
Syntax
ACQuire?
Examples
ACQUIRE? might return the following string for the current acquisition
parameters:
ACQUIRE:CURRENTCOUNT:ACQWFM S 0;HISTWFMS
0;HISTHITS 0;MASKWFMS 0;MASKSAMPLES 0;MASKTHITS 0;MASKHITS1
0;MASKHITS2 0;MASKHITS3 0;MASKHITS4 0;MASKHITS5 0;MASKHITS6
0;MASKHITS7 0;MASKHITS8 0;:ACQUIRE:MODE SAMPLE;NUMAVG
16;SAVEFILE:SAVESCREEN "";SAVEWFM "";:ACQUIRE:STATE
1;STOPAFTER:MODE RUNSTOP;BELL 0;ACTION NO NE;CONDITION
ACQWFMS;COUNT 1
.
ACQuire:CURRentcount:ACQWfms? (Query Only)
This query only command returns the current count value of acquired waveforms.
The target value of this count is set by the ACQuire:STOPAfter:COUNt command
(in conjunction with the ACQuire:STOPAfter:CONDition command). The
instrument then counts up to this value. When the count reaches (or exceeds) the
value, acquisition stops, and the specied StopAfter action is enabled.
Group
Acquisition
Syntax
ACQuire:CURRentcount:ACQWf ms?
Related Commands
ACQuire:STOPAfter:COUNt, ACQuire:STOPAfter:CONDition
Returns
NR1 is the current count value of acquired waveforms.
Examples
ACQUIRE:CURRENTCOUNT:ACQWF MS? might return
ACQUIRE:CURRENTCOUNT:ACQWF MS 20, indicating that currently 20 waveforms
have been acquired.
DSA/CSA/TDS8X00/B Series Programmer Manual 2-51

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DSA8200 Series and is the answer not in the manual?

Tektronix DSA8200 Series Specifications

General IconGeneral
BrandTektronix
ModelDSA8200 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals