Commands Listed in Alphabetical Order
ACQuire:STOP
After:CONDition
This command sets or queries the StopAfter condition. The StopAfter condition
qualifies a stop condition for the acquisition system. Only one StopAfter condition
can be a ctive
at a g iven time. Each StopAfter condition identifies, directly or
indirectly, a specific data e lement or operation such that all mutually exclusive
conditions are unique and unambiguous. This command allows you to specify
the condition on which to stop acquiring. The condition is valid when the
ACQuire:STOPAfter:MODE is set to CONDition. This is equivalent to checking
the Condition control in the Stop After section of the Acquisition Setup dialog
box and se
lecting a condition.
NOTE. You must have the associated feature enabled before you can
set the StopAfter condition. For example, you must turn on histograms
(HIStogram:ENABle) before you can set the HISTHits StopAfter condition.
Group
Acquisition
Syntax
ACQuire:STOPAfter:CONDiti on { ACQWfms | HISTWaveform
| HISTH its | MASKW aveform | MASKSample | MASK<x>Hits |
MASKTOTalhit | FRAMecy cle | AVGComp }
ACQuire:STOPAfter:CONDiti on?
Related Commands
ACQuire:STOPAfter:COUNt, ACQuire:STOPAfter:MODe
Arguments
ACQWfms sets the instrument to stop acquiring after some specified number of
raw acquisition cycles. This setting tells the instrument to count the number
of Main Timebase sweeps (Mag sweeps are not counted independently) and
stop acquisition after the s pecifi ed number of acquisitions has been reached.
Use the ACQuire:STOPAfter:COUNt command to set the target number of
waveforms.
HISTWaveform sets the instrument to stop acquisition after a
specified number of Histogram source waveforms is acquired. Use the
ACQuire:STOPAfter:COUNt command to set the specified number of
waveforms.
2-62 DSA/CSA/TDS8X00/B Series Programmer Manual