Commands Listed in Alphabetical Order
ACQuire:STOP
After:COUNt
This command sets or queries the target StopAfter count for the condition
specified by the ACQuire:STOPAfter:CONDition command. The current count
for the condi
tion must be equal to or greater than this value before acquisitions
are stopped and a StopAfter action is enabled. The state of the numeric
StopAfter count for each condition is kept individually so that you do not need
to re-enter a count when switching between conditions. Use the appropriate
ACQuire:CURRentcount command to get the current count for a condition (see
Related Commands below).
Group
Acquisition
Syntax
ACQuire:STOPAfter:COUNt <NR1>
ACQuire:STOPAfter:COUNt?
Related Commands
ACQuire:STOPAfter:MODe, ACQuire:STOPAfter:CONDition,
ACQuire:CURRentcount:ACQWfms?, ACQuire:CURRentcount:HISTWfms?,
ACQuire:CURRentcount:HISTHits?, ACQuire:CURRentcount:MASKWfms?,
ACQuire:CURRentcount:MASKHits<x>?, ACQuire:CURRentcount:
MASKSamples?
Arguments
NR1 is the count value that must be reached (or exceeded) before the acquisitions
stop and StopAfter action can occur.
Examples
ACQUIRE:STOPAFTER:COUNT 12 sets the StopAfter count for the specified
condition to 12.
ACQUIRE:STOPAFTER:COUNT? might return ACQuire:STOPAfter:COU Nt 5,
indicating that the total count for the specified condition is 5.
ACQuire:STOPAfter:MODe
This command tells the instrument when to stop taking acquisitions. The
query form of this command returns the StopAfter mode. This command is the
equivalent of checking Run/Stop or Condition in the Stop After section of the
Acquisition Setup dialog box. (The ACQuire:STATE command can also be used
to start or stop acquisitions.)
Group
Acqui
sition
2-64 DSA/CSA/TDS8X00/B Series Programmer Manual