Commands Listed in Alphabetical Order
Group
Cursor
Synta
x
CURSor:WAVeform?
Examples
CURSOR:WAVEFORM? might return CURSOR:WAVEFORM:HDELTA
0.487243652344;VDELTA -19.9398797595E-009 ; HPOS1
0.241729736328;HPOS2 -0.24551391601 6; POSITION1
31.1302605210E-009; POSITION2 51.0701402806E-009
.
CURSor:WAVeform:HDELTa? (Query Only)
This query-only command returns the horizontal difference between the waveform
cursors. This is the absolute value of the horizontal position of the first cursor
minus the horizontal position of the second cursor. This is equivalent to the
waveform Dt readout value.
Group
Cursor
Syntax
CURSor:WAVeform:HDELTa?
Examples
CURSOR:WAVEFORM:HDELTA? might return CU RSOR:WAVEFORM:HDELTA
1.03E-9
, indicating that the time between the waveform cursors is 1.03 ns.
CURSor:WAVeform:HPOS<x>? (Query Only)
This query only command returns the position of the specified waveform cursor.
The cursor is specified by x, which can be 1 or 2. This corresponds to the v1 or v2
(cursor 1 or cursor 2) cursor readout.
Group
Cursor
Syntax
CURSor:WAVeform:HPOS<x>?
Examples
CURSOR:WAVEFORM:HPOS1? might return CURSOR:WAVEFORM: HPOS2
0.247344970703
, indicating that cursor 2 is at 247.3 mV with respect to ground
on the source waveform.
2-106 DSA/CSA/TDS8X00/B Series Programmer Manual