EasyManua.ls Logo

Tektronix DSA8200 Series

Tektronix DSA8200 Series
431 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Getting Started
To change eithe
r of the GPIB settings, do the following:
1. Select User Preferences from the Utilities menu.
2. Select the GPIB Conguration Tab.
3. Change the G PIB Address to a unique address.
4. Clic
kOK.
The instrument is now set up for bidirectional communication with your controller.
Mas
ter/Slave Mode
The DSA8200 is factory set to be a GPIB device (slave), allowing you to control
the instrument via the GPIB port. You c an change the instrument to b e a GPIB
controller (master), allowing you to control other devices via the GPIB port.
GPIB/VISA Controller/Device (Master/Slave) Switch. The DSA8200 can be
congured as a G PIB Device (Slave) or Controller (Master). The factory setting
is congured to enable control of the instrument via the GPIB port, referred to
as the “Device Mode”.
DSA/CSA/TDS8X00/B Series Programmer Manual 1-5

Related product manuals