Commands Listed in Alphabetical Order
HORizontal:F
RAMescan:RESET (No Query Form)
This command (no query form) resets and restarts the FrameScan acquisition at
the specified start bit. Sending this command is equivalent to clicking the Reset
buttoninthe
FrameScan section of the Horizontal Setup dialog box. You can set
the start bit w ith the HORizontal:FRAMescan:STARTBit command.
Group
Horizontal
Syntax
HORizontal:FRAMescan:RESE T
Examples
HORIZONTAL:FRAMESCAN:RESE T resets and restarts the FrameScan acquisition
at the specified start bit.
HORizontal:FRAMescan:SCANBits
This command sets or queries the number of bits or subframes to s can in the
FrameScan m ode. Sending this command is equivalent to setting a value in the
Scan Bits control in the FrameScan section o f the Horizontal Setup dialog box.
Scanning will begin at either the start bit specified with the
HORizontal:FRAMescan:STARTBit command or the current horizontal bit
location. The minimum value of bits to scan is 1 and the maximum value is 2
20
-1.
Group
Horizontal
Syntax
HORizontal:FRAMescan:SCAN Bits <NR1>
HORizontal:FRAMescan:SCAN Bits?
Related Commands
HORizontal:FRAMescan:STARTBit
Arguments
NR1 sets the number of bits to be scanned in FrameScan mode. The range is 1
through 1,048,575 bits.
Examples
HORIZONTAL:FRAMESCAN:SCAN BITS 4000 sets the number of bits scanned in
the FrameSc
an mode to 4000.
HORIZONTAL:FRAMESCAN:SCAN BITS? might return
HORIZONTA
L:FRAMESCAN:SCANBITS 700
, indicating the number of s can
bits is set to 700.
2-162 DSA/CSA/TDS8X00/B Series Programmer Manual