Commands Listed in Alphabetical Order
Group
Horizontal
Syntax
HORizontal:MAGnify<x>:REF Point <NR2>
HORizontal:MAGnify<x>:REF Point?
Related Commands
HORizontal:MAGnify<x>:SCAle
Arguments
NR2 is the fraction of the record at which the horizontal reference point is set. The
range is 0 through 1 (corresponding to 0% through 100% of record); the default
is 0.5. The resolution is 0.01.
Examples
HORIZONTAL:MAG1:REFPOINT 2.5E-1 sets the horizontal reference p
oint
for the Mag1 timebase to 0.25 (25%). This is the point around which the
HORizontal:MAG:SCAle command contracts or expands the waveform.
HORIZONTAL:MAG2:REFPOINT? might return
HORIZONTAL:MAGNIFY2:REFPO INT .500000000000, indicating that the
horizontal reference point for the Mag2 timebase is set to 0.5 (50%), which
is the default.
HORizontal:MAGnify<x>:RESolution? (Query Only)
This query only c ommand returns the current resolution per s
ample of the
specified Mag timebase. The Mag timebase is specified by x, which can be 1 or
2. The resolution for the selected timebase is displayed in the Resolution field in
the Timebase section of the Horizontal Setup dialog box. You cannot directly set
the resolution. Instead, use the HORizontal:MAG<x>:SCAle command to set the
horizontal scale and the HORizontal:MAG<x>:RECORDLength command to set
the record length. Resolution is displayed in horizonta
l units, typically time.
Group
Horizontal
Syntax
HORizontal:MAGnify<x>:RES olution?
Related Commands
HORizontal:MAGnify<x>:SCAle, HORizontal:MAGnify<x>:RECordlength
Examples
HORIZONTAL:MAG1:RESOLUTIO N? might return
HORIZONTAL:MAGNIFY1:RESOL UTION 6.31000000000E–012, indicating
that the resolution is 6.31 ps.
2-168 DSA/CSA/TDS8X00/B Series Programmer Manual