EasyManuals Logo

Tektronix DSA8200 Series User Manual

Tektronix DSA8200 Series
431 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #191 background imageLoading...
Page #191 background image
Commands Listed in Alphabetical Order
HORizontal:M
AIn:SCAle
This command sets or q ueries the scale (time p er division) for the Main timebase.
Sending this command is equivalent to selecting MainTB and adjusting the Scale
control in th
e Timebase section of the Horizontal Setup dialog box.
The horizontal scale that you set with this command is common to all channel
waveforms u
sing the Main timebase. The horizontal scale, together with the
record length setting, determines the sample interval.
Group
Horizont
al
Syntax
HORizontal:MAIn:SCAle <NR3>
HORizontal:MAIn:SCAle?
Related Commands
HORizontal:MAIn:RECordlength, HORizontal:MAIn:RESolution?
Arguments
NR3 is the horizontal time per division for the Main timebase. The range is 1 ps
per division to 5 ms per division.
Examples
HORIZONTAL:MAIN:SCALE 2.0E-9 sets the main scale to 2 ns per division.
HORIZONTAL:MAIN:SCALE? might return HORIZONTAL:MAIN:S CALE
10.0000000000E-009
, indicating that the Main timebase is currently set to
10 ns per division.
HORizontal:MAIn:TOFPoint? (Query Only)
This query only command returns time of rst point, which is the time interval
between the trigger point and the rst point in a Main timebase record (waveform).
Refer to the illustration below for a diagram of the relationships of time of rst
point, time of last point, horizontal position, and horizontal reference point.
DSA/CSA/TDS8X00/B Series Programmer Manual 2-177

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DSA8200 Series and is the answer not in the manual?

Tektronix DSA8200 Series Specifications

General IconGeneral
BrandTektronix
ModelDSA8200 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals