Commands Listed in Alphabetical Order
Syntax
MEASUrement:M
EAS<x>:REFLevel<x>:ABSolute: LOW <NR3>
MEASUrement:MEAS<x>:REFLev el<x>:ABSolute:LOW?
Related Commands
MEASUrement:MEAS<x>:REFLevel<x>:METHod
Arguments
NR3 is the low reference level in absolute user units. The default is 0 V.
Examples
MEASUREMENT:MEAS3:REFLEVEL 1:ABSOLUTE:LOW 2.0E-2 sets the low
reference level for Measurement 3 to 20 mV.
MEASUREMENT:MEAS2:REFLEVEL 1:ABSOLUTE:LOW? might return
MEASUREMENT:MEAS2:REFLEVEL 1:ABSOLUTE:LOW 3.00000000000E-0 03,
indicating that the low reference level for Measurement 2 to is se t to 3 mV.
MEASUrement:MEAS<x>:REFLevel<x>:ABSolute:MID
This command sets or queries the mid reference level in absolute user units for
the specified measurement. The measurement slot is specified by x, which can
be 1 through 8. The reference level is specified by x (which corresponds to the
Source1 or Source2 waveform) and can be 1 or 2. (You can set two different mid
reference levels when taking measurements, such as phase or delay, between
two source waveforms.)
This absolute value is used as the mid reference level when
MEASUrement:MEAS<x>:REFLevel<x>:METHod is set to ABSolute. Sending
this command is equivalent to selec ting A BSolute in the Reference Level Calc
Method section of the RefLevel dialog and then entering an absolute value for
Mid in the Reference section of the Measurement Setup dialog box.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:REFLev el<x>:ABSolute:MID <NR3>
MEASUrement:MEAS<x>:REFLev el<x>:ABSolute:MID?
Related Commands
MEASUrement:MEAS<x>:REFLevel<x>:METHod
Arguments
NR3 is the mid reference level in absolute user units. The default is 0 V.
Examples
MEASUREMENT:MEAS3:REFLEVEL 2:ABSOLUTE:MID 4.0E–2 sets the mid
reference level for Measurement 3 on Source 2 to 40 mV.
DSA/CSA/TDS8X00/B Series Programmer Manual 2-231