Commands Listed in Alphabetical Order
TRIGGER:AUTOS
YNC:PLENGTH?
might return TRIGGER:AUT OSYNC:PLENGTH
0
, indicating that AutoSync is not set to sync the pattern length setting when it
executes.
TRIGger:CH<x>:CLKRec:LIST? (Query Only)
This query returns a list of the clock recovery selections that are available for
the channel specified by x, which can be 1 through 4. The available choices are
dependent on the optical sampling module that is installed.
The optical modules have differing valid settings. Use the <>:LIST? query to
determine the valid {Clock Recovery | Filter | Bandwidth | Wavelength} settings
for your optical module. Features of optic al modules are listed in their respective
user manuals. Features of currently installed optical modules can be viewed u sing
the System Properties dialog.
Group
Trigger
Syntax
TRIGger:CH<x>:CLKRec:LIST ?
Examples
TRIGGER:CH1:CLKREC:LIST? might return TRI GGER:CH1:CLKREC:LIST
NONE,OC12,OC48
.
TRIGger:CH<x>:CLKRec:RANge
Sets or queries the allowable range for a User (custom) clock rate as <nr3>-<nr3>
for the channel specified by X, which can be 1 through 4. (The available
choices for x are dependent on the sampling module that is installed.) The
channel specified must be from an sampling module that provides continuous
clock-recovery capability in order to return a range; otherwise, "NONE" is
returned.
The allowable range is the range s
upported by the module providing the specified
channel; values within the range returned by this query c an be used to specify the
User clock rate when u sing the TRIGger:CH<x>:CLKRec:USEr command.
Group
Trigger
Syntax
TRIGger:CH<x>:CLKRec:RANg e
TRIGger:CH<x>:CLKRec:RANg e?
Related Commands
TRIGger:CH<x>:CLKRec:USEr, SYSTem:PROPerties:CH<x>:CLKUser?
2-304 DSA/CSA/TDS8X00/B Series Programmer Manual