Status and Events
Table 3-7: Syst
em Event Messages
Code Message
400
Query event
401
Power on (PON bit 7 set)
402
Operation complete (OPC bit 0 set)
403
User request (URQ bit 6 set)
404
Power fail (DDE bit 3 set)
405 Request co
ntrol
410
Query INT
ERRUPTED (Q YE bit 2 set)
420
Query UN
TERMINATED (QYE bit 2 set)
430
Query D E
ADLOCKED (Q YE bit 2 set)
440
Query U
NTERMINATED after indenite response (QYE bit 2 s et)
Execution Warning
The fo
llowing table lists warning messages that do not interrupt the ow of
command execution. These notify you that you may get unexpected results.
Table 3-8: Execution Warning Messages (EXE Bit 4)
Code Message
500 Execution warning
510
String data too long, truncated
525 Par
ameter underrange
526 Parameter overrange
527 Parameter rounded
52
8
Parameter out of range
5
30
D
ata stop > start. Values swapped internally
531
D
ata stop > record length, Curve truncated
532
Curve data too long, Curve truncated
540 Measurement warning
541 Measurement warning, Low signal amplitude
542 Measurement warning, Unstable histogram
543 Measurement warning, Low resolution
544 Measurement warning, Uncertain edge
545 Measurement warning, Invalid minmax
546 Measurement warning, Need 3 edges
547
Measurement warning, Clipping positive/negative
548
Measurement warning, Clipping positive
549
Measurement warning, Clipping negative
3-18 DSA/CSA/TDS8X00/B Series Programmer Manual

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DSA8200 Series and is the answer not in the manual?

Tektronix DSA8200 Series Specifications

General IconGeneral
BrandTektronix
ModelDSA8200 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals