Commands Listed in Alphabetical Order
Group
Calibration
Syntax
CALibrate:DATE:MAInframe?
Examples
CALIBRATE:DATE:MAINFRAME? might return CALIBRATE:DATE:MAINFRAME
15 JAN 00 16:25
, indicating that the last calibration update for the mainframe
wasdoneonJ
anuary 15, 2000, at 4:25 PM.
CALibrate:DCCALibrator
This command sets or queries the value of the DC Calibrator voltage.
Group
Calibration
Syntax
CALibrate:DCCALibrator <NR2>
CALibrate:DCCALibrator?
Arguments
NR2 is the value to which you want to set the DC Calibrator voltage. The range
for this value is from -1.25 VDC through +1.25 VDC.
Examples
CALIBRATE:DCCALIBRATOR 0.5 sets the DC Calibrator voltage to 0.5 V.
CALIBRATE:DCCALIBRATOR? might return CALIBRATE:DCCALIBRATO R
5.00000000000E-001
, indicating that the DC Calibrator voltage is set to 0.5 V.
CALibrate:HOSTinfo:CH<x>? (Query Only)
This query only command returns the mainframe channel, model number, and
serial number in which the sampling module channel was located during the last
calibration information update.
Group
Calibration
Syntax
CALibrate:HOSTinfo:CH<x>?
Examples
CALIBRATE:HOSTINFO:CH2 might return CALIBRATE:HOSTI NFO:CH2
"DSA8200,B010235,CH4"
, indicating that the sampling module currently in
channel 2 was installed in channel 4 of the mainframe, serial number B010235,
during the last calibration information update.
DSA/CSA/TDS8X00/B Series Programmer Manual 2-73