Commands Listed in Alphabetical Order
CALibrate:LO
CK:STATus? (Query Only)
This is a query only command that returns the status of the calibration protection
mode.
Group
Calibration
Syntax
CALibrate:LOCK:STATus?
Examples
CALIBRATE:LOCK:STATUS? might return CALIBRATE:LOCK:STATUS 1,
indicating that the calibration information is protected and cannot be updated.
CALibrate:STATus:CH<x>? (Query Only)
This is a query only command that returns the current calibration status for the
sampling module channel.
Group
Calibration
Syntax
CALibrate:STATus:CH<x>?
Returns
May return PASS, DEFAULTS, or WARMUP.
Examples
CALIBRATE:STATUS:CH1? might return CALIBRATE:STAT US:CH1 PASS,
indicating that the calibration status of the sampling module in channel 1 is PASS.
CALibrate:STATus:MAInframe? (Query Only)
This is a query only command that returns the current calibration status for the
mainframe.
Group
Calibration
Syntax
CALibrate:STATus:MAInfram e?
Returns
May return PASS, DEFAULTS, or WARMUP.
2-74 DSA/CSA/TDS8X00/B Series Programmer Manual

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DSA8200 Series and is the answer not in the manual?

Tektronix DSA8200 Series Specifications

General IconGeneral
BrandTektronix
ModelDSA8200 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals