Commands Listed in Alphabetical Order
Examples
CALIBRATE:STA
TUS:MAINFRAME?
might return
CALIBRATE:STATUS:MAINFRAME PASS, indicating that the calibration status
of the m ainframe is PASS.
CALibrate:T
EMPerature:CH<x>? (Query Only)
This is a query only command that returns the difference (in °C) between the
current sampling module channel temperature and the temperature recorded at
the last ca
libration information update.
Group
Calibration
Syntax
CALibrate:TEMPerature:CH<x >?
Examples
CALIBRATE:TEMPERATURE:CH2? might return
CALIBRATE:TEMPERATURE:CH2 1.5, indicating that the temperature difference
between the current sampling module channel temperature and the temperature
recorded at the last calibration information update is 1.5 °C.
CALibrate:TEMPerature:MAInframe? (Query Only)
This is a query only command that returns the difference (in °C) between the
current mainframe temperature and the temperature recorded at the last calibration
information update.
Group
Calibration
Syntax
CALibrate:TEMPerature:MAIn frame?
Examples
CALIBRATE:TEMPERATURE:MAIN FRAME? might return
CALIBRATE:TEMPERATURE:MAIN FRAME -2.7, indicating that the
difference in temperature between the current mainframe temperature and the
temperature recorded at the last calibration information update is -2.7 °C
CALibrate:UPDATEinfo:ALL (No Query Form)
This command (no query form) updates the calibration information for the
mainframe and all resident module channels into nonvolatile memories.
DSA/CSA/TDS8X00/B Series Programmer Manual 2-75